表題 | Safety of the batteries and power units used in insulin pumps: A pilot cross-sectional study by the Association for the Study of Innovative Diabetes Treatment in Japan |
掲載誌名 | 正式名:Journal of Diabetes Investigation 略 称:JDI |
掲載区分 | 国外 |
巻・号・頁 | pp.online |
著者・共著者 | Murata Takashi,Nirengi Shinsuke ,Sakane Naoki,Kuroda Akio, Hirota Yushi,Matsuhisa Munehide ,Namba Mitsuyoshi,Kobayashi Tetsuro ,and for The Association for the Study of Innovative Diabetes Treatment in Japan (ASINDTJ) |
発行年月 | 2017/11 |
概要 | Abstract
Aims/Introduction We investigated the safety of the batteries and power units used in insulin pumps in Japan. Materials and Methods A self-administered questionnaire was sent to the 201 members of the Association for Innovative Diabetes Treatment in Japan. Results A total of 56 members responded, and among the 1,499 active devices, 66 had episodes of trouble related to the batteries and power units. The ratio of reported troubles to the number of insulin pumps was significantly higher in insulin pumps with a continuous glucose monitoring sensor compared with insulin pumps without a continuous glucose monitoring sensor (odds ratio 2.82, P < 0.05). The cause and the consequences varied. The brands of the batteries varied; alkaline batteries purchased at drug stores and other shops accounted for 19.7%. Termination of battery life within 72 h of use was reported most frequently (50.0%), suspension of the insulin pump (21.2%) and leakage of the battery fluid (4.5%) followed. A total of 53.2% of the reported insulin pumps needed to be replaced, and 37.1% of them recovered after replacement of the battery. Conclusions As trouble related to the batteries and power units of insulin pumps was frequent, practical guidance should be provided to respective patients regarding the use of reliable batteries, and to be well prepared for unexpected insulin pump failure. |
DOI | 10.1111/jdi.12765 |