表題 | Simulation analysis of three intubating supraglottic devices during infant chest compression |
掲載誌名 | 正式名:Pediatrics International |
巻・号・頁 | 2015(57),pp.180-2 |
著者・共著者 | Kohama Hanako, Komasawa Nobuyasu, Ueki Ryusuke, Kaminoh Yoshiroh, Nishi Shin-ichi |
発行年月 | 2015 |
概要 | Abstract
Current guidelines for pediatric cardiopulmonary resuscitation suggest that supraglottic devices are alternatives for tracheal intubation with minimal interruption of chest compression. We examined the utility of three intubating supraglottic devices, air-Q® (air-Q), Ambu® aura-i (aura-i), and i-gel® (i-gel), utilizing manikin simulation. Twenty-two novice physicians performed securing of airway on an infant manikin with the three devices. We measured the rate of success on ventilation and the insertion time with or without chest compression. Successful ventilation rate did not significantly decrease with chest compression in the three devices (without chest compression: air-Q, 21/22; aura-i, 20/22; i-gel, 20/22, during chest compression: air-Q, 20/22; aura-i, 20/22; i-gel, 18/22). The insertion time with air-Q and aura-i did not extend significantly for chest compression. In contrast, the insertion time with i-gel was significantly extended in chest compression (P < 0.05). Air-Q and aura-i are more useful for airway management during chest compression than i-gel. |